Perform a Control Solution Test:
•BeforetestingwiththeGLUCOCARD™01-minibloodglucosemeter
for the first time.
•Whenyouopenanewbottleofteststrips.
•Wheneveryoususpectthemeterorteststripsmaynotbe
functioning properly.
•Iftestresultsappeartobeabnormallyhighorloworarenot
consistent with clinical symptoms.
•Theteststripbottlehasbeenleftopenorhasbeenexposedto
temperatures below 34°F (1°C) or above 86°F (30°C), or humidity
levels above 80%.
•Tocheckyourtechnique.
•WhentheGLUCOCARD™01-minibloodglucosemeterhasbeen
dropped or stored below 32°F (0°C) or above 122°F (50°C).
Important:
•UseonlywithGLUCOCARD™01-minibloodglucosemeterand
GLUCOCARD™ 01-SENSOR blood glucose test strips.