Arkray Glucocard X-Meter Manuale delle Istruzioni Pagina 48

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 72
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 47
48
Before You Compare
Perform a control solution test to make sure your meter and test strips are
working correctly. Wait at least 2 hours after a meal or drink (other than water)
before doing comparison blood tests. Fasting is essential if venous blood is
required for the reference analyzer.
When at the Laboratory
The capillary samples for both tests must be taken and tested within 15
minutes of each other. The sample must be collected with anticoagulant
appropriate to the laboratory analyzer. The recommended technique should
be followed. Analysis should be performed following the manufacturer’s
guidelines.
ALWAYS wash hands in warm, soapy water. Rinse and dry them thoroughly
before testing.
ALWAYS use fresh capillary whole blood on meter tests.
Vedere la pagina 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 71 72

Commenti su questo manuale

Nessun commento