50
IMPORTANT!
Insert strip
VERTICALLY
into blood sample,
NOT at an angle.
c) Press the release button on the lancing device to lance
the skin.
Continue to hold the lancing device firmly
against the skin until a blood drop forms.
Step 4:
Pick up the meter and touch the end of the test strip to the blood sample until
the reaction site is full. Immediately remove the meter and test strip from the
blood drop.
IMPORTANT
•Repeatblooddrawifuidisnotadeepred
color.
•DONOTTESTBLOODifsampleissmeared,
flattened out, or not forming a round drop.
Repeat steps starting with Step 3.
•Ifittakeslongerthan20SECONDStoobtain
a blood sample, and any sample has been
applied to the strip, discard strip and use a
new strip before proceeding.